LINEARITY TEST ANALYSIS OF AIRCRAFT RADIATION OUTPUT X-RAY FLUOROSCOPY

  • Bambang Harianto
    Gunadarma University, Central Jakarta, Indonesia
  • Muhammad Kardaji
    Gunadarma University, Central Jakarta, Indonesia
Issue Vol. 1 No. 4 (2023): MSJ-December
Release 29-12-2023
Section Articles
Pages 1-5
Abstract

The number or intensity of X-ray photons produced correlates with the increase in tube current per time (mAs), which affects the image density on the final film. To ensure the consistency of the increase in X-ray radiation output and the Linear Coefficient of X-ray radiation output, is within the tolerance limits specified in PERKA BAPETEN Regulation Number 9 of 2011. The aim of this research is to determine the linearity of the X-ray Fluoroscopy radiation output in one of the houses. Jakarta sick. The method used is an experimental approach with a quantitative model. Measurement of radiation output using a PM1610 Personal Dosimeter on a Fluoroscopy aircraft. The results of measuring the linearity of X-ray radiation output with voltage variations of 40 kV, 50 kV, 60 kV, 70 kV show that the value of the Linearity Coefficient (CL) is (0.08 ; 0.06 ; 0.09 ; 0.07). So it can be said that the X-Ray Fluoroscopy equipment is in good condition because the CL value obtained in this study is below the Tolerance Limit Value according to PERKA BAPETEN Number 9 of 2011, namely CL ≤ 0.1 or ≤10%. This research can be carried out as a reference in further research

Keywords
Linearity Radiation Output X-Rays Fluoroscopy
How to Cite (APA)
Harianto, B., & Kardaji, M. (2023). LINEARITY TEST ANALYSIS OF AIRCRAFT RADIATION OUTPUT X-RAY FLUOROSCOPY. MSJ : Majority Science Journal, 1(4), 1–5. https://doi.org/10.61942/msj.v1i4.50
Licensing & Copyright

Downloads

Download data is not yet available.
References
  1. Aizah, H. S. (2023). Pengaruh faktor eksposi terhadap dosis radiasi dan kualitas citra pesawat radiografi umum pada pemodelan phantom thoraks (Doctoral dissertation, Universitas Islam Negeri Maulana Malik Ibrahim).
  2. Akhadi, M. (2020). Sinar-X Menjawab Masalah Kesehatan. Sleman : Deepublish CV. Budi Utama.
  3. Anggraeni, L. (2020). Faktor-Faktor Yang Berhubungan Dengan Penggunaan Pocket Dosemeter Pada Petugas Radiologi di Instalasi Radiologi 9 Rumah Sakit B Andar Lampung. Jurnal Teras Kesehatan, 3(1), 70-85.
  4. Dewilza, N., Rad, A. M., Kes, M. T., Yudha, S., & Kes, M. T. (2023). Dasar-Dasar CT-Scan. Sleman : Deepublish CV. Budi Utama.
  5. Hidayah, N., Astari, F. M., Rad, T. R., & Fakhrurreza, M. (2021). Uji Kebocoran Tabung Pesawat Sinar-X Mobile Di Universitas ‘Aisyiyah Yogyakarta (Doctoral dissertation, Universitas ‘Aisyiyah Yogyakarta).
  6. Hyperastuty, A. S., Mukhammad, Y., & Sugeng, S. (2021). Analisis Uji Kesesuaian Pesawat Sinar X Radiografi Mobile Merk Drgem Topaz-40d Menggunakan X-Ray Multimeter PIRANHA. Journal Of Health Science (Jurnal Ilmu Kesehatan), 6(1), 19-26.
  7. Indah, NH, Dewang, S., & Astuty, S.D. (2023). Analisis Laju Dosis Keluaran Pesawat Sinar-X Fluoroskopi Dual Fungsi Di Rs Wahidin Sudirohusodo Makassar. Berkala Fisika , 26 (1), 8-14. https://ejournal.undip.ac.id/index.php/berkala_fisika/article/view/54922
  8. Irsal, M., A. Targian., S. E. Wulandari., N. Kinasih. (2023). Perbandingan Dosis Radiasi Pemeriksaan Radiografi Thoraks PA Menggunakan Faktor Eksposi: Standar dan High kVp Berdasarkan Database DRL Si-INTAN BAPETEN. Jurnal Pengawasan Tenaga Nuklir Volume 3, Nomor 2.
  9. Khoirot, R. M., Hakim, M. H., Alam, Y., & Ananingtyas, R. S. A. (2023). Analisis Nilai Drl Parameter Esak/Inak Pemeriksaan Thorax Ap/Pa X-Ray Canon. Journal of Science Nusantara, 3(2), 61-68.
  10. Nuramdiani, D., Sarengat., I. Maulana. (2021). Tinjauan Densitas Radiograf Pada Berbagai Ketebalan Step Wedge Berbasis Variasi Arus Tabung. Jurnal Pendidikan Fisika Tadulako Online. Volume 9 No. 2.
  11. Nurcahyo, P., & Aryani, A. I. (2023). Pelaksanaan Quality Control Pesawat Sinar X di Pelayanan Radiodiagnostik. JRI (Jurnal Radiografer Indonesia), 6(2), 61-64. https://doi.org/10.55451/jri.v6i2.152
  12. Rani, R. (2020). Uji Akurasi Alat Ukur Radiasi Pada Kasus Kebocoran Tabung Pesawat Mobile X-Ray di BPFK Makassar (Doctoral dissertation, Universitas Hasanuddin).
  13. Surachman. (2023). “Metode Estimasi Risiko Kanker Pada Organ Berisiko Pemeriksaan Radiografi Kepala di RSUD Kabupaten Bekasi Menggunakan Metode Monte Carlo”. Skripsi. Jakarta : Universitas Nasional.
  14. Wiharja, U., & Al Bahar, A. K. (2019). Analisa Uji Kesesuaian Pesawat Sinar-X Radiografi. Prosiding Semnastek. https://jurnal.umj.ac.id/index.php/semnastek
  15. Yufita, E., & Safitri, R. (2023). Analisa Pengaruh Faktor Eksposi Pesawat Sinar-X Terhadap Densitas Optik Film Radiografi. Jurnal Hadron, 5(01), 9-14.